Talk with one of our sales

Monday to Friday 8AM - 5PM WIB

021- 72791867

Product

Home / Product

WIKA

CPC6050 -- Modular Pressure Controller

Category: PRESSURE, CALIBRATORS

Model: Pressure Controller, Pressure Controllers

Avaibility: OUT OF STOCK

Call For Price

Design

The highly configurable model CPC6050 modular pressure controller offers maximum flexibility to best suit the customers’ requirements. The instrument can have up to two independent pressure regulating channels which can operate simultaneously. Each channel can have up to two sensors. The instrument can also have an optional barometric reference for gauge or absolute pressure emulation. This instrument can be specified as a desktop or as 19" rack-mounting kit.

Application

The controller offers many applications within calibration laboratories and production environments because of its pressure range -1 … 210 bar (-15 … 3,045 psi) and accuracy down to 0.01 % IS-50. Its ability to control pressures as low as 25 mbar (10" of H20) with a high stability makes it the ideal calibration and verification solution for healthcare and aerospace industries.

Simultaneous calibration channels along with interchangeable plug-and-play pressure sensors and an intuitive GUI makes CPC6050 an easy-to-use and maintain instrument.

Functionality

The touchscreen, along with an intuitive user interface, provide maximum ease-of-use. The large number of menu languages add to its operability. In addition to specifying a certain pressure set point either by entering it via touchscreen or sending it via remote interface, the pressure can be changed in defined, programmable step sizes by using the STEP buttons. Moreover, the user can also easily create extensive test programs using the instrument menu. Depending on the application, the rate of control can be either pre-set precision, high speed or a user-defined variable rate.

Software

The WIKA-Cal calibration software enables the convenient calibration of pressure measuring instruments and the generation of test certificates. Additionally, the instrument can also be remotely controlled using the serial command formats, the Mensor standard, SCPI or further optional command sets are available.

Complete test and calibration systems

On request, complete mobile or stationary test systems can be manufactured. There is an IEEE-488.2, RS-232, USB and an Ethernet interface for communication with other instruments, and thus the instrument can be integrated into existing systems.

Backward compatibility

Pressure range

Standard

Optional

Accuracy 1)

0.01 % FS 2)

0.01 % IS-50 3)

Gauge pressure

0 ...

0.025 to 0

... 210 bar

0 ...

1 to 0 ...

210 bar

(0 ...

0.36 to 0

... 3,045 psi)

(0 ...

15 to 0 ... 3,045 psi)

Bi-directional

-0.012

... 0.012 to -1

... 210 bar

-1 ...

10 to -1

... 210 bar

Absolute pressure 4)

(-0.18

... 0.18 to -15

... 3,045 psi)

(-15

... 145 to -15 ... 3,045 psi)

0 ...

0.5 to 0 ... 211 bar abs.

0 ...

1 to 0 ... 211 bar abs.

Precision 5)

(0 ...

7.5 to 0 ... 3,060 psi abs.)

(0 ...

15 to 0 ... 3,060 psi abs.)

0.004 % FS

0.004 % FS

Calibration interval

365 days 6)

365 days

Optional barometric reference

The barometric reference can be used to switch pressure types 7), absolute <=> gauge. With gauge

Function

pressure sensors, the measuring range of the sensors must begin with -1 bar (-15 psi) in order to carry

out an absolute pressure emulation.

Measuring range

552

...

1,172 mbar abs. (8 ... 17 psi abs.)

Accuracy 1)

0.01 % of reading

Pressure units

39 and two freely programmable

The highly configurable CPC6050 can also be used with pressure sensors of its predecessor model CPC6000. The sensors can be used individually or together with the CPR6050, hence providing the user a complete backward capability.

Pressure range

Standard

Optional

Accuracy 1)

0.01 % FS 2)

0.01 % IS-50 3)

Gauge pressure

0 ...

0.025 to 0

... 210 bar

0 ...

1 to 0 ...

210 bar

(0 ...

0.36 to 0

... 3,045 psi)

(0 ...

15 to 0 ... 3,045 psi)

Bi-directional

-0.012

... 0.012 to -1

... 210 bar

-1 ...

10 to -1

... 210 bar

Absolute pressure 4)

(-0.18

... 0.18 to -15

... 3,045 psi)

(-15

... 145 to -15 ... 3,045 psi)

0 ...

0.5 to 0 ... 211 bar abs.

0 ...

1 to 0 ... 211 bar abs.

Precision 5)

(0 ...

7.5 to 0 ... 3,060 psi abs.)

(0 ...

15 to 0 ... 3,060 psi abs.)

0.004 % FS

0.004 % FS

Calibration interval

365 days 6)

365 days

Optional barometric reference

The barometric reference can be used to switch pressure types 7), absolute <=> gauge. With gauge

Function

pressure sensors, the measuring range of the sensors must begin with -1 bar (-15 psi) in order to carry

out an absolute pressure emulation.

Measuring range

552

...

1,172 mbar abs. (8 ... 17 psi abs.)

Accuracy 1)

0.01 % of reading

Pressure units

39 and two freely programmable

Screen

8.9" colour LC display with resistive touchscreen

Resolution

4 ... 6 digits depending on range and units

Connections

Pressure connections

up to 8 ports with 7/16"- 20 F SAE, up to 2 ports with 1/8" F NPT and 1 port with 10-32 UNF female

Filter elements

The instrument has 40-micron filters on all pressure ports.

Pressure port adapters

Standard: without

Option: 6 mm tube fitting, 1/4" tube fitting, 1/4" female NPT fittings, 1/8" female NPT fittings or 1/8"

female BSP fittings

Barometer port adapters

Standard: barb fitting

Option: 6 mm tube fitting, 1/4" tube fitting

Permissible pressure media

Dry, clean air or nitrogen (ISO 8573-1:2010 class 5.5.4 or better)

Wetted parts

Aluminium, brass, 316 and 316L stainless steel, Buna N, FKM/FPM, PCTFE, PEEK, PTFE, PPS,

glass-filled epoxy, RTV, ceramic, silicone, silicone grease, urethane

Overpressure protection

Safety relief valve fixed to reference pressure sensor and adjusted to specific customised

measuring range

Permissible pressure

Supply port

~ 110 % FS

Measure/Control port

max. 105 % FS

Voltage supply

Power supply

AC 100 ... 240 V, 50 Hz / 60 Hz

Power consumption

max. 120 VA

Permissible ambient conditions

Storage temperature

0 ... 70 °C (32 ... 158 °F)

Humidity

5 … 95 % r. h. (non-condensing)

Compensated temperature range

15 ... 45 °C (59 ... 113 °F)

Mounting position

horizontal

Control parameter

SVR module 8)

LPPump module

Control stability

< 0>

Control mode

precision, high speed and customised

external supply on/off 9)

Control time

15 s 10)

25 s 10)

Control range

0 ... 100 % FS

Minimum control pressure

0.0017 bar (0.025 psi) over exhaust pressure or

0.0034 bar (0.05 psi) over exhaust pressure or

0.05 % FS whichever is greater

0.05 % FS whichever is greater

Overshoots

< 0>

< 0>

(typical < 0>

(< 0>

Test volume

50 ... 1,000 ccm

50 ... 300 ccm

Communication

Interface

Standard: Ethernet, IEEE-488, USB, RS-232

Option: WiFi (with a USB-WiFi adapter)

Command sets

Mensor, WIKA SCPI, others optional

Response time

approx. 100 ms

Internal program

up to 24 sequences with up to 99 steps each

OTHER PRODUCT

We Offer Many Product For All Your Needs

Need help buying the right product?

We’ll assist you for further information, application assistance, or to schedule a technical meeting.