Design
The highly configurable model CPC6050 modular pressure controller offers maximum flexibility to best suit the customers’ requirements. The instrument can have up to two independent pressure regulating channels which can operate simultaneously. Each channel can have up to two sensors. The instrument can also have an optional barometric reference for gauge or absolute pressure emulation. This instrument can be specified as a desktop or as 19" rack-mounting kit.
Application
The controller offers many applications within calibration laboratories and production environments because of its pressure range -1 … 210 bar (-15 … 3,045 psi) and accuracy down to 0.01 % IS-50. Its ability to control pressures as low as 25 mbar (10" of H20) with a high stability makes it the ideal calibration and verification solution for healthcare and aerospace industries.
Simultaneous calibration channels along with interchangeable plug-and-play pressure sensors and an intuitive GUI makes CPC6050 an easy-to-use and maintain instrument.
Functionality
The touchscreen, along with an intuitive user interface, provide maximum ease-of-use. The large number of menu languages add to its operability. In addition to specifying a certain pressure set point either by entering it via touchscreen or sending it via remote interface, the pressure can be changed in defined, programmable step sizes by using the STEP buttons. Moreover, the user can also easily create extensive test programs using the instrument menu. Depending on the application, the rate of control can be either pre-set precision, high speed or a user-defined variable rate.
Software
The WIKA-Cal calibration software enables the convenient calibration of pressure measuring instruments and the generation of test certificates. Additionally, the instrument can also be remotely controlled using the serial command formats, the Mensor standard, SCPI or further optional command sets are available.
Complete test and calibration systems
On request, complete mobile or stationary test systems can be manufactured. There is an IEEE-488.2, RS-232, USB and an Ethernet interface for communication with other instruments, and thus the instrument can be integrated into existing systems.
Backward compatibility
Pressure range |
Standard |
|
|
Optional |
|
|||
Accuracy 1) |
0.01 % FS 2) |
|
|
0.01 % IS-50 3) |
||||
Gauge pressure |
0 ... |
0.025 to 0 |
... 210 bar |
0 ... |
1 to 0 ... |
210 bar |
||
|
(0 ... |
0.36 to 0 |
... 3,045 psi) |
(0 ... |
15 to 0 ... 3,045 psi) |
|||
Bi-directional |
-0.012 |
... 0.012 to -1 |
... 210 bar |
-1 ... |
10 to -1 |
... 210 bar |
||
Absolute pressure 4) |
(-0.18 |
... 0.18 to -15 |
... 3,045 psi) |
(-15 |
... 145 to -15 ... 3,045 psi) |
|||
0 ... |
0.5 to 0 ... 211 bar abs. |
0 ... |
1 to 0 ... 211 bar abs. |
|||||
Precision 5) |
(0 ... |
7.5 to 0 ... 3,060 psi abs.) |
(0 ... |
15 to 0 ... 3,060 psi abs.) |
||||
0.004 % FS |
|
|
0.004 % FS |
|
||||
Calibration interval |
365 days 6) |
|
|
365 days |
|
|||
Optional barometric reference |
The barometric reference can be used to switch pressure types 7), absolute <=> gauge. With gauge |
|||||||
Function |
||||||||
|
pressure sensors, the measuring range of the sensors must begin with -1 bar (-15 psi) in order to carry |
|||||||
|
out an absolute pressure emulation. |
|
|
|
||||
Measuring range |
552 |
... |
1,172 mbar abs. (8 ... 17 psi abs.) |
|
|
|
||
Accuracy 1) |
0.01 % of reading |
|
|
|
|
|||
Pressure units |
39 and two freely programmable |
|
|
|
The highly configurable CPC6050 can also be used with pressure sensors of its predecessor model CPC6000. The sensors can be used individually or together with the CPR6050, hence providing the user a complete backward capability.
Pressure range |
Standard |
|
|
Optional |
|
|||
Accuracy 1) |
0.01 % FS 2) |
|
|
0.01 % IS-50 3) |
||||
Gauge pressure |
0 ... |
0.025 to 0 |
... 210 bar |
0 ... |
1 to 0 ... |
210 bar |
||
|
(0 ... |
0.36 to 0 |
... 3,045 psi) |
(0 ... |
15 to 0 ... 3,045 psi) |
|||
Bi-directional |
-0.012 |
... 0.012 to -1 |
... 210 bar |
-1 ... |
10 to -1 |
... 210 bar |
||
Absolute pressure 4) |
(-0.18 |
... 0.18 to -15 |
... 3,045 psi) |
(-15 |
... 145 to -15 ... 3,045 psi) |
|||
0 ... |
0.5 to 0 ... 211 bar abs. |
0 ... |
1 to 0 ... 211 bar abs. |
|||||
Precision 5) |
(0 ... |
7.5 to 0 ... 3,060 psi abs.) |
(0 ... |
15 to 0 ... 3,060 psi abs.) |
||||
0.004 % FS |
|
|
0.004 % FS |
|
||||
Calibration interval |
365 days 6) |
|
|
365 days |
|
|||
Optional barometric reference |
The barometric reference can be used to switch pressure types 7), absolute <=> gauge. With gauge |
|||||||
Function |
||||||||
|
pressure sensors, the measuring range of the sensors must begin with -1 bar (-15 psi) in order to carry |
|||||||
|
out an absolute pressure emulation. |
|
|
|
||||
Measuring range |
552 |
... |
1,172 mbar abs. (8 ... 17 psi abs.) |
|
|
|
||
Accuracy 1) |
0.01 % of reading |
|
|
|
|
|||
Pressure units |
39 and two freely programmable |
|
|
|
Screen |
8.9" colour LC display with resistive touchscreen |
|
Resolution |
4 ... 6 digits depending on range and units |
|
Connections |
|
|
Pressure connections |
up to 8 ports with 7/16"- 20 F SAE, up to 2 ports with 1/8" F NPT and 1 port with 10-32 UNF female |
|
Filter elements |
The instrument has 40-micron filters on all pressure ports. |
|
Pressure port adapters |
Standard: without |
|
|
Option: 6 mm tube fitting, 1/4" tube fitting, 1/4" female NPT fittings, 1/8" female NPT fittings or 1/8" |
|
|
female BSP fittings |
|
Barometer port adapters |
Standard: barb fitting |
|
|
Option: 6 mm tube fitting, 1/4" tube fitting |
|
Permissible pressure media |
Dry, clean air or nitrogen (ISO 8573-1:2010 class 5.5.4 or better) |
|
Wetted parts |
Aluminium, brass, 316 and 316L stainless steel, Buna N, FKM/FPM, PCTFE, PEEK, PTFE, PPS, |
|
|
glass-filled epoxy, RTV, ceramic, silicone, silicone grease, urethane |
|
Overpressure protection |
Safety relief valve fixed to reference pressure sensor and adjusted to specific customised |
|
|
measuring range |
|
Permissible pressure |
|
|
Supply port |
~ 110 % FS |
|
Measure/Control port |
max. 105 % FS |
|
Voltage supply |
|
|
Power supply |
AC 100 ... 240 V, 50 Hz / 60 Hz |
|
Power consumption |
max. 120 VA |
|
Permissible ambient conditions |
|
|
Storage temperature |
0 ... 70 °C (32 ... 158 °F) |
|
Humidity |
5 … 95 % r. h. (non-condensing) |
|
Compensated temperature range |
15 ... 45 °C (59 ... 113 °F) |
|
Mounting position |
horizontal |
|
Control parameter |
SVR module 8) |
LPPump module |
Control stability |
< 0> |
|
Control mode |
precision, high speed and customised |
external supply on/off 9) |
Control time |
15 s 10) |
25 s 10) |
Control range |
0 ... 100 % FS |
|
Minimum control pressure |
0.0017 bar (0.025 psi) over exhaust pressure or |
0.0034 bar (0.05 psi) over exhaust pressure or |
|
0.05 % FS whichever is greater |
0.05 % FS whichever is greater |
Overshoots |
< 0> |
< 0> |
|
(typical < 0> |
(< 0> |
Test volume |
50 ... 1,000 ccm |
50 ... 300 ccm |
Communication |
|
|
Interface |
Standard: Ethernet, IEEE-488, USB, RS-232 |
|
|
Option: WiFi (with a USB-WiFi adapter) |
|
Command sets |
Mensor, WIKA SCPI, others optional |
|
Response time |
approx. 100 ms |
|
Internal program |
up to 24 sequences with up to 99 steps each |
|