Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
Measurement location-independent operation and Bluetooth measurement value display via Smartphone/tablet and
testo 330i App
Non-paper documentation and report creation directly on site with the testo 330i App
Reliable flue gas probe attachment with the testoFix probe mount
Robust, completely sealed plastic housing for use in tough conditions
The testo 330i is the combination of proven technology and revolutionary handling. The basis of the flue gas analyzer
is the measurement technology of the proven predecessor testo 330 LL, with these outstanding features:
The operation of the measuring instrument and the display of the measurement values take place by Bluetooth via the testo 330i App on your Smartphone/tablet. This means you always have all relevant values at your fingertips right where you need them. After finishing the measurement, you can insert comments or photos of the plant into your report, and send it to your customer or your office by e-mail. Today's way to measure flue gas.
General technical data
Compatability |
requires iOS 7.1 or newer / Android 4.3 or newer |
requires mobile end device with Bluetooth 4.0 |
|
Storage temperature |
-20 to +50 °C |
Operating temperature |
-5 to +45 °C |
Power supply |
Rech. batt. block 3.7 V / 2.6 Ah |
Mains unit 6 V / 1.2 A (optional) |
|
Memory |
500000 readings |
Weight |
720 g (excluding battery) |
Dimensions |
270 x 160 x 57 mm |
Guarantee
Guarantee conditions |
Instrument/probe/gas sensors O2, CO: 48 months Gas sensor NO: 24 months Thermocouple and rech. battery: 12 months https://www.testo.com/guarantee |
Measurement parameter |
Measuring range |
Accuracy (± 1 digit) |
Resolution |
Temperature (dependent on the thermocouple in the flue gas probe) |
-40 to +1200 °C |
±0.5 °C (0.0 - +100.0 °C) ±0.5 % of m.v. (remaining meas. range) |
±0.1 °C (-40 - +999.9 °C) ±1 °C (remaining meas. range) |
Draught measurement |
-9.99 to +40 hPa |
±0.02 hPa or ± 5 % of m.v. (-0.50 to +0.60 hPa) ±0.03 hPa (+0.61 to +3.00 hPa) ±1.5 % of m.v. (+3.01 to +40.00 hPa) (the greater value applies) |
0.01 hPa |
Pressure measurement |
0 to 300 hPa |
±0.5 hPa (0.0 to +50.0 hPa) ±1 % of m.v. (+50.1 to +100.0 hPa) ±1.5 % of m.v. (remaining meas. range) |
0.1 hPa |
O2 measurement |
0 to 21 vol.% |
±0.2 vol.% |
0.1 vol.% |
CO measurement (not H2-compensated) |
0 to 4000 ppm |
±20 ppm or ±10 % of m.v. (0 to 400 ppm) ±5 % of m.v. (401 to 2000 ppm) ±10 % of m.v. (2001 to 4000 ppm) |
1 ppm |
CO measurement (not H2-compensated) with activated measuring range extension |
0 to 15000 ppm |
±200 ppm or ±20 % of m.v. (0 to 15000 ppm) |
1 ppm |
CO measurement (H2-compensated) |
0 to 8000 ppm |
±10 ppm or ±10 % of m.v. (0 to 200 ppm) ±20 ppm or ±5 % of m.v. (201 to 2000 ppm) ±10 % of m.v. (2001 to 8000 ppm) |
1 ppm |
CO measurement (H2-compensated) with activated measuring range extension |
0 to 30000 ppm |
±200 ppm or ±20 % of m.v. (0 to 30000 ppm) |
1 ppm |
Efficiency testing (Eta) |
0 to 120 % |
0.1 % |
|
Flue gas loss |
0 to 99.9 % |
0.1 % |
|
CO2 determination (Calculation from O2) |
Display area 0 to CO2 max |
±0.2 vol.% |
0.1 vol.% |
Option: NO measurement: |
0 to 3000 ppm |
±5 ppm (0 to 100 ppm) ±5 % of m.v. (101 to 2000 ppm) ±10 % of m.v. (2001 to 3000 ppm) |
1 ppm |