TC101A -- Thermocouple Data Logger
Category: TEMPERATURE
Model: Temperature Data Logger
Avaibility: AVAILABLE
Call For Price
Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
TC101A -- Thermocouple Data Logger
Category: TEMPERATURE
Model: Temperature Data Logger
Avaibility: AVAILABLE
Call For Price
The TC101A temperature data logger, is a compact versatile device to be used with thermocouple probes for accurate temperature monitoring and profiling. This data logger is an ideal and affordable solution for applications such as HVAC studies, oven profiling, dry-ice shipments, process validation and many general purpose temperature monitoring applications.
Compatible with eight different types of thermocouple probes, the TC101A can measure temperatures as low as -270 °C , and up to as high as 1820 °C (probe dependent), with an environmental operating temperature range of -40 °C to +80 °C for the data logger body.
The 10 year battery life and the memory storage capacity of 500,000 readings with optional memory wrap makes the TC101A ideal for extended deployments. With the user friendly MadgeTech Data Logger Software, the TC101A is easily connected to a PC to download the collected data. The MadgeTech Software is capable of presenting the data in graphic, tabular, and summary formats for reporting. Data can also be exported easily to Excel for further analysis and calculations.
Internal Channel |
|||
Temperature Range: |
-40 °C to +80 °C (-40 °F to +176 °F |
||
Temperature Resolution: |
0.01 °C (0.018 °F) |
||
Calibrated Accuracy: |
±0.5 °C (±0.9 °F |
||
Remote Channels |
|||
Thermocouple Connection : |
|
||
Cold Junction Compensation : |
Automatic based on internal channel |
||
Max. Thermocouple Resistance : |
100 ? |
||
Thermocouple Type |
Range (ºC) |
Resolution |
Accuracy |
J |
-210 to +760 |
0.1 ºC |
±0.5 ºC |
K |
-270 to +1370 |
0.1 ºC |
±0.5 ºC |
T |
-270 to +400 |
0.1 ºC |
±0.5 ºC |
E |
-270 to +980 |
0.1 ºC |
±0.5 ºC |
R |
-50 to +1760 |
0.5 ºC |
±2.0 º C |
S |
-50 to +1760 |
0.5 ºC |
±2.0 º C |
B |
+50 to +1820 |
0.5 ºC |
±2.0 º C |
N |
-270 to +1300 |
0.1 ºC |
±0.5 ºC |
*Thermocouple accuracy is specified with a 24 AWG |
|||
Reading Rate : |
1 reading every second up to 1 reading every 24 hours |
||
Memory: |
|
||
Wrap Around : |
Yes |
||
Start Modes : |
|
||
Stop Modes : |
|
||
Multiple Start/Stop Mode : |
Start and stop the device multiple times without having to download data or communicate with a PC |
||
Multiple Start/Stop Mode Activation : |
To start the device: Press and hold the pushbutton for 5 seconds, the green LED will flash during this time. The device has started logging. To stop the device : Press and hold the pushbutton for 5 seconds, the red LED will flash during this time. The device has stopped logging. |
||
Real Time Recording : |
The device may be used with PC to monitor and record data in real time. |
||
Alarm : |
Alarm: Programmable high and low limits; alarm is activated when temperature reaches or exceeds sets limits. Alarm Delay: A cumulative alarm delay may be set in which the device will activate the alarm (via LED) only when the device has recorded a user specified time duration of data. |
||
Trigger Settings : |
High and Low limits may be set for the thermocouple channel. Once data meets or exceed sets limits, the device will record to memory. Bi-level start and stop triggers can also be programmed. Users can specify the number of readings to take after the device triggers. |
||
LED Functionality : |
Green LED blinks: 10 second rate to indicate logging, 15 second rate to indicate delay start mode Red LED blinks: 10 second rate to indicate low battery and/or full memory, 1 second rate to indicate an alarm condition |
||
Password Protection : |
An optional password may be programmed into the device to restrict access to configuration options. Data may be read out without the password. |
||
Battery Type : |
3.6V lithium battery included; user replaceable 10 years typical at a 15 minute reading rate |
||
Data Format : |
Date and time stamped °C, °F, K, °R; µV, mV, V |
||
Time Accuracy: |
±1 minute/month (at 20 ºC, stand alone data logging) |
||
Computer Interface: |
USB (interface cable required); 115,200 baud |
||
Software: |
XP SP3/Vista/Windows 7/Windows 8 |
||
Operating Environment |
-40 °C to +80 ºC (-40 °F to +176 °F), 0 %RH to 95 %RH non- condensing |
||
Dimensions: |
1.4 in x 2.2 in x 0.6 in (36 mm x 56 mm x 16 mm)
|
||
Weight: |
0.9 oz (24 g) |
||
Approvals: |
CE |