Hi Temp 140 PT -- High Temperature Data Logger
Category: DATA LOGGER
Model: Temperature
Avaibility: AVAILABLE
Call For Price
Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
Hi Temp 140 PT -- High Temperature Data Logger
Category: DATA LOGGER
Model: Temperature
Avaibility: AVAILABLE
Call For Price
The HiTemp140-PT is a submersible, temperature data logger that can operate up to 140 °C (284 °F) and has an accuracy of +/-0.1 °C (0.18 °F). The HiTemp140-PT comes in two different probe lengths, both probe styles feature a 24 inch bendable stainless steel probe. The HiTemp140-PT-1 has a 1 inch probe tip and where the HiTemp140-PT-5 has a 5 inch probe tip. Both probes can measure extended temperatures up to 350 °C (662 °F) The probe is durable and can be spiraled, bent or angled in any direction, making it easy to log temperatures within bottles, vials or other hard to reach places. The HiTemp140-PT records date and time stamped readings, and has non-volatile solid state memory which retains data even if the battery becomes discharged.
Using the latest MadgeTech Software, starting, stopping and downloading the HiTemp140-PT is simple and easy. To use, simply place the HiTemp140-PT in the IFC400 or IFC406 docking station
(sold separately). Using the software, an immediate or delay start can be chosen, as well as the reading rate. Start the data logger, remove it from the docking station and the device is ready to be deployed. Graphical, tabular and summary data is provided for analysis and data can be viewed in °C, °F, K or °R. The data can also be automatically exported to Excel® for further calculations.
Temperature Sensor: |
100 ? Platinum RTD |
|
|
|
|
Probe Measurement |
-200 °C to +350 °C (-328 °F to +662 °F) |
|
Range: |
*BODY OF LOGGER CANNOT EXCEED 140 °C |
|
|
|
|
Temperature |
0.01 °C (0.02 °F) |
|
Resolution: |
||
|
||
|
|
|
|
±0.1 °C/±0.18 °F |
|
|
(20 °C to +140 °C/68 °F to +284 °F) |
|
Calibrated Accuracy: |
±0.3 °C/±0.54 °F |
|
(-20 °C to +19.99 °C/-4 °F to +67.98 °F) |
||
|
||
|
±0.4 °C/±0.72 °F |
|
|
(-40 °C to –20.01 °C/-40 °F to -4.02 °F) |
|
|
|
|
Start Modes: |
• Software programmable immediate start |
|
• Delay start up to 18 months in advance |
||
|
||
|
|
|
Stop Modes: |
Manual or Timed (specific date and time) |
|
|
|
|
|
May be used with PC to monitor and record data |
Real Time Recording: |
in real time |
||||||||||||
|
|||||||||||||
|
|
||||||||||||
|
An optional password may be programmed into the device |
||||||||||||
Password Protection: |
to restrict access to configuration options. Data may be |
||||||||||||
|
read out without the password. |
||||||||||||
|
|
||||||||||||
Memory: |
32,700 readings |
||||||||||||
|
|
||||||||||||
Wrap Around: |
Yes |
||||||||||||
|
|
||||||||||||
Reading Rate: |
1 reading every second up to 1 reading every 24 hours |
||||||||||||
|
|
||||||||||||
|
3.6V high-temperature lithium battery included;
|