CPH6000 -- Process Calibrator
Category: PRESSURE, CALIBRATORS
Model: Portable Calibrators, Portable Calibrator
Avaibility: OUT OF STOCK
Call For Price
Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
CPH6000 -- Process Calibrator
Category: PRESSURE, CALIBRATORS
Model: Portable Calibrators, Portable Calibrator
Avaibility: OUT OF STOCK
Call For Price
Handling
For a solution that always matches the application, there are many pressure sensors to choose from, with accuracies of up to 0.025 % and measuring ranges up to 6,000 bar (87,000 psi), which can be interchanged quickly and without tools. In addition to being fixed to the instrument, the reference pressure sensor can, optionally, be used externally via an approx. 1.2 m (4 ft) long connecting cable. If the reference pressure sensor on the instrument is changed, then the digital indicator will recognise the new measuring range and so this saves the operator from having to configure it via the menu.
Functionality
In the setup menu there are 3 operating modes to choose from: MEASURING, CALIBRATION and PRESSURE SWITCH-TEST. In the CALIBRATION and PRESSURE SWITCH-TEST modes, a menu assistant supports the user during each operation and records, for example, the calibration data from several calibrations or automatically calculates the switch hysteresis. To power the test item and to read its measurement signals, there are electrical inputs and outputs which are protected from adverse conditions in the field by captive protection caps.
Software
For the evaluation and documentation of the calibration data stored in the CPH6000, WIKA-Cal calibration software is available. Using this software, the data is automatically transferred into a printable calibration certificate. Furthermore, WIKA-Cal also offers, over and above PC-supported calibration, the management of the calibration and instrument data in an SQL database. For data transfer, an RS-232 and a USB interface are available.
Complete test and service cases
For maintenance and service applications, various service case systems are available. These include service cases with or without pressure generation, charger, test-cable set, adapter, etc.
Certified accuracy
For each reference pressure sensor, the accuracy is certified by a factory calibration certificate which accompanies the instrument. On request, we can provide a DKD/DAkkS calibration certificate for this instrument.
|
Sensor technology |
1 reference pressure sensor (interchangeable without tools) 1); |
|
|
|
|
|
|
||||||||
|
|
Option: external operation via 1.2 m (4 ft) cable |
|
|
|
|
|
|
|
|
||||||
|
Measuring range |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Gauge |
bar |
-0.25 ... |
+0.25 |
-0.4 |
... +0.4 |
-0.6 |
... +0.6 |
-1 |
... 0 |
|
-1 ... |
1.5 |
|
||
|
|
|
-1 ... |
2.5 |
|
-1 ... |
3 |
|
-1 ... |
5 |
-1 |
... 9 |
|
-1 ... |
15 |
|
|
|
|
-1 ... |
24 |
|
-1 ... |
39 |
|
0 ... |
0.25 |
0 ... |
0.4 |
|
0 ... |
0.6 |
|
|
|
|
0 ... |
1 |
|
0 ... |
1.6 |
|
0 ... |
2.5 |
0 ... |
4 |
|
0 ... |
6 |
|
|
|
|
0 ... |
10 |
|
0 ... |
16 |
|
0 ... |
25 |
0 ... |
40 |
|
0 ... |
60 |
|
|
|
|
0 ... |
100 |
|
0 ... |
160 |
|
0 ... |
250 |
0 ... |
400 |
|
0 ... |
600 |
|
|
|
|
0 ... |
1,000 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
psi |
-4 ... |
+4 |
|
-6 ... |
+6 |
|
-10 ... |
+10 |
-14.5 ... |
0 |
-14.5 ... |
22 |
||
|
|
|
-14.5 ... |
36 |
-14.5 ... |
44 |
-14.5 ... 73 |
-14.5 ... |
130 |
-14.5 ... |
220 |
|||||
|
|
|
-14.5 ... |
350 |
-14.5 ... |
565 |
0 ... |
4 |
0 ... |
6 |
|
0 ... |
10 |
|
||
|
|
|
0 ... |
14,5 |
0 ... |
25 |
|
0 ... |
40 |
0 ... |
60 |
|
0 ... |
90 |
|
|
|
|
|
0 ... |
145 |
|
0 ... |
250 |
|
0 ... |
360 |
0 ... |
580 |
|
0 ... |
870 |
|
|
|
|
0 ... |
1,450 |
0 ... |
2,320 |
0 ... |
3,630 |
0 ... |
5,800 |
0 ... |
8,700 |
||||
|
|
|
0 ... |
14,500 |
|
|
|
|
|
|
|
|
|
|
|
|
|
Accuracy of the measuring chain |
0.025 % FS 2) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Gauge |
bar |
0 ... |
1,600 |
0 ... |
2,500 |
0 ... |
4,000 |
0 ... |
5,000 |
0 ... |
6,000 |
||||
|
|
psi |
0 ... |
23,200 |
0 ... |
36,200 |
0 ... |
58,000 |
0 ... |
72,500 |
0 ... |
87,000 |
||||
|
Accuracy of the measuring chain |
0.1 % FS 2) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Absolute pressure |
bar abs. |
-0.8 |
... 1.2 |
0 ... |
0.25 |
0 ... |
0.4 |
0 ... |
0.6 |
|
0 ... |
1 |
|
||
|
|
|
0 ... |
1.6 |
|
0 ... |
2.5 |
|
0 ... |
4 |
0 ... |
6 |
|
0 ... |
10 |
|
|
|
|
0 ... |
16 |
|
0 ... |
25 |
|
|
|
|
|
|
|
|
|
|
|
psi abs. |
-11.6 ... |
17.4 |
0 ... |
4 |
|
0 ... |
6 |
0 ... |
8 |
|
0 ... |
14.5 |
||
|
|
|
0 ... |
23 |
|
0 ... |
36 |
|
0 ... |
60 |
0 ... |
90 |
|
0 ... |
145 |
|
|
|
|
0 ... |
230 |
|
0 ... |
360 |
|
|
|
|
|
|
|
|
|
|
Accuracy of the measuring chain |
0.025 % FS 2) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Overpressure limit |
3 times; < 25> |
|
|
|
|
3 times; < 360> |
... ? 8,700 psi |
|
|
|
|||||
|
(dependent on the measuring |
2 times; > 25 bar ... ? 600 bar |
|
|
|
2 times; > 360 psi |
|
|
|
|||||||
|
range) |
1,5 times; > 600 bar |
... ? 2,500 bar |
|
1,5 times; > 8,700 psi ... |
? 36,260 psi |
|
|
||||||||
|
|
1,2 times; > 2,500 bar |
|
|
|
1,2 times; > 36,260 psi |
|
|
|
|
||||||
|
Types of pressure |
Relative pressure, {absolute pressure from 0 ... |
25 bar abs. (0 ... 360 psi abs.) and vacuum from |
|||||||||||||
|
|
-1 ... 39 bar (-14.5 ... |
550 psi)} |
|
|
|
|
|
|
|
|
|
|
|
||
|
Sensor compatibility |
Compatible with model CPT6000 reference pressure sensors |
|
|
|
|
|
|
||||||||
OTHER PRODUCT