Talk with one of our sales

Monday to Friday 8AM - 5PM WIB

021- 72791867

Product

Home / Product

WIKA

CPH6510 -- Intrinsically Safe Hand-Held Pressure Calibrator

Category: PRESSURE, CALIBRATORS

Model: Portable Calibrators, Portable Calibrator

Avaibility: OUT OF STOCK

Call For Price

General information

The operator can choose between numerous different pressure measuring ranges. The model CPH65I0 can have up to two different reference pressure sensors integrated within it. These reference pressure sensors are fixed into the housing. The pressure connection is found on the bottom of the calibrator.

Various application possibilities

With up to two reference pressure sensors integrated into the CPH65I0, combined with current input, pressure switch function and a resistance thermometer, the CPH65I0 enables the calibration of practically any pressure instrument. The additional ATEX approval expands the possible applications of this calibrator into hazardous areas.

Accuracy

The CPH65I0 offers an accuracy of up to 0.025 % of the span in 24 pressure measuring ranges. The measurements can be displayed in one of 16 standard units.

Certified accuracy

For each CPH65I0 reference pressure sensor, the accuracy is certified by a factory calibration certificate which accompanies the instrument. On request, we can provide a DKD/DAkkS calibration certificate for this instrument.

Measuring range

mbar

-25 ... +25 1)

-70 … +70 1)

-350 ... +350 1)

-500 … +500 1)

Overpressure limit

mbar

70

200

700

1,000

Accuracy

% FS

0.1

0.05

0.035

Measuring range

bar

-1 ... +1 1)

-1 ... +2 1)

Overpressure limit

bar

2

4

Accuracy

% FS

0.025

Measuring range

bar

0 … 1 1)

0

... 2 1)

0 ... 3.5 2)

0 ... 7 2)

0 ... 10 2)

0 ... 20 2)

Overpressure limit

bar

2

4

13

13

13

40

Accuracy

% FS

0.025

Measuring range

bar

0 ... 35 2)

0

... 70

0 ... 100

0 ... 200

0 ... 350

0 ... 700

Overpressure limit

bar

70

200

200

400

700

1,000

Accuracy

% FS

0.025

0.035

Measuring range

bar abs.

0 ... 1

0

... 2

0 ... 7

0 ... 10

0 ... 20

Overpressure limit

bar abs.

2

4

13

13

40

Accuracy

% FS

0.025

Measuring range

mbar diff.

0 ... 25 1) 3)

0

... 70 1) 3)

0 ... 350 1) 3)

0 ... 2,000 1) 3)

0 ... 3,500 1) 3)

0 ... 7,000 1) 3)

Overpressure limit

mbar diff.

70

200

700

4,000

7,000

10,000

Accuracy

% FS

0.1

0.05

0.035

0.025

Type of pressure

Relative pressure, absolute pressure, vacuum and differential pressure 4)

Pressure connection

? NPT female (incl. adapter 1/8 NPT male to G ½ B male) 5)

Pressure medium

all liquids and gases which are compatible with 316 SS stainless steel 1)

Resolution

5-digit

Current

Measuring range

0 … 24 mA

Resolution

1 µA

Accuracy

0.015 % of measured value ±2 µA

Temperature

Measuring range

-40 … +150 °C

Resolution

0.01 °C

Accuracy

0.015 % of measured value ±20 m?, or 0.2 °C for complete measuring chain

(Pt100 resistance thermometer and CPH65I0)