TC 100 -- Precision Thermocouple Calibrator
Category: TEMPERATURE
Model: Handheld Temperature Calibrator and Indicator
Avaibility: AVAILABLE
Call For Price
Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
TC 100 -- Precision Thermocouple Calibrator
Category: TEMPERATURE
Model: Handheld Temperature Calibrator and Indicator
Avaibility: AVAILABLE
Call For Price
The TC-100 Thermocouple Calibrator provides high accuracy source and measurement of ten common thermocouples, as well as mV. The TC-100’s accuracy of ±0.3°C for Type J T/Cs includes all errors, at resolutions of ±0.01°C or °F in measure mode and ±0.1°C or °F in source mode. Features including MIN/MAX re-call in measure mode, three setpoints per thermocouple range, a large knob for decade control of the output in source mode, and the ability to accept bare T/C wires in addition to mini-plug inputs, makes the TC-100 an accurate, easy-to-use instrument for all your thermocouple calibration needs.
|
unless otherwise noted) |
Input Voltage |
|
Range |
-10 to +75.000 mV |
Resolution |
1 µV |
Accuracy |
±0.007 % of rdg, ±10 µV |
Input Impedance |
> 1 MegOhm |
Output Voltage |
|
Range |
-10 to +75.000 mV |
Resolution |
1 µV |
Accuracy |
±0.007 % of rdg, ±10 µV |
Output Impedance |
> 1 Ohm |
Thermocouple Source/Measure |
|
Types |
J, K, T, E, R, S, B, L, U, C |
Range |
mV |
Resolution |
|
Source |
±0.1°C or °F |
Measure |
±0.01°C or °F |
Accuracy |
|
J |
±0.5 °C; -210 °C to -100 °C |
|
±0.3 °C; -100 °C to +1,200 °C |
K |
±0.6 °C; -200 °C to -100 °C |
|
±0.35 °C; -100 °C to +1,000 °C |
|
±0.5 °C; +1,000 °C to +1,372 °C |
T |
±0.7 °C; -200 °C to -150 °C |
|
±0.3 °C; -150 °C to +400 °C |
E |
±0.5 °C; -200 °C to -100 °C |
|
±0.3 °C; -100 °C to +1,000 °C |
R |
±1.8 °C; 0 °C to 250 °C |
|
±1.0 °C; 250 °C to +1,767 °C |
S |
±1.8 °C; 0 °C to 250 °C |
|
±1.0 °C; 250 °C to +1,767 °C |
B |
±1.7 °C; 600 °C to 1,000 °C |
|
±1.2 °C; 1,000 °C to 1,820 °C |
L |
±0.5 °C; -200 °C to -100 °C |
|
±0.4 °C; -100 °C to +900 °C |
U |
±0.7 °C; -200 °C to 0 °C |
|
±0.3 °C; 0 °C to +600 °C |
C |
±0.4 °C; 0 to °C 1,000 °C |
|
±0.7 °C; 1,000 °C to +1,800 °C |
|
±1.2 °C; +1,800 °C to +2,316 °C |
CJC Temp. Offset |
±0.05 °C/°C outside of 23±5 °C |
Warm-up Time |
1 minute to specification |
Environmental |
|
Operating Temperature |
-10 °C to +55 °C |
Storage Temperature |
-20 °C to 70 °C |
|
|
Power Requirements |
9 VDC |
Battery |
9 V alkaline; 006P/ IEC 6F22/ |
|
NEDA1604 |
|
Optional NiCad |
|
Optional AC adapter/charger |
|
|
Mechanical |
|
Dimensions |
5.7” H x 3.15” W x 1.43” D |
|
(144.7 x 80.0 x 36.3 mm) |
Weight |
12 ounces (340 grams) |
|
|
Optional Accessories |
Carrying Case;Model LCA-05A |
Notes: |
|