Testo Saveris 2
Category: DATA LOGGER, TEMPERATURE, HUMIDITY
Model: Humidity, Temperature, Temperature Data Logger, Humidity Data Logger
Avaibility: AVAILABLE
Call For Price
Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
Testo Saveris 2
Category: DATA LOGGER, TEMPERATURE, HUMIDITY
Model: Humidity, Temperature, Temperature Data Logger, Humidity Data Logger
Avaibility: AVAILABLE
Call For Price
The testo Saveris 2 WiFi data logger system is the modern solution for monitoring temperature and humidity values as well as CO2 concentration in storerooms and work rooms. Installing the system is child's play, and can be carried out either via a browser, or with the testo Saveris 2 App. The WiFi data loggers reliably record temperature, humidity and CO? concentration at adjustable intervals and transmit the measurement values directly by wireless LAN to the Testo Cloud.
The readings stored can be analyzed at any time, anywhere, using an internet-enabled smartphone, tablet or PC.Limit value violations are immediately reported by e-mail, optionally by SMS or via the testo Saveris 2 App as a push notification. This allows critical processes to be kept always under control, even if you are not on site. The long battery life additionally ensures that the testo Saveris 2 system needs to be serviced only rarely.
Our packages
The Testo Cloud is the core operating element for setting up your testo Saveris 2 system. Here you can configure your WiFi data loggers, set limit value alarms and analyze your measurement data. You must first register at www.saveris.net to have access to the Testo Cloud. Depending on the desired range of functions, you have a choice between the free Basic and more extensive Advanced functionality for use of the Testo Cloud. In both packages, you have access to an API interface, in order to export measurement data to your systems.
Basic |
Advanced |
|||
Measuring cycle |
15 min to 24 h |
1 min to 24 h |
||
Communication cycle |
15 min to 24 h |
1 min to 24 h |
||
Data storage |
Max. 3 months |
Max. 2 years |
||
Reports |
Manual (.pdf/.csv) |
Manual (.pdf/.csv) Automatic (.pdf/.csv) |
||
Data analysis |
For one measurement site each (external probes count as measurement sites) |
For up to 10 measurement channels simultaneously |
||
Number of users per account |
1 |
10 |
||
Number of WiFi data loggers per account |
Unlimited |
Unlimited |
||
Alarm options |
Upper/lower alarm limits |
|
||
System notifications |
|
|
||
E-mail alarm |
Yes |
Yes |
||
SMS alarm |
No |
|
||
12-month licence Order no. 0526 0735 |
24-month licence Order no. 0526 0732 |
36-month licence Order no. 0526 0733 |
Technical data
testo Saveris 2- T1 |
testo Saveris 2- T2 |
testo Saveris 2- T3 |
testo Saveris 2- H1 |
testo Saveris 2- H2 |
testo 160 IAQ |
|||
Temperature measurement |
||||||||
Sensor type |
NTC internal |
NTC |
TC Type K |
TC Type J |
TC type T |
NTC internal |
NTC |
|
Measuring range |
-30 to +50 °C |
-50 to +150 °C |
-195 to +1350 °C |
-100 to +750 °C |
-200 to +400 °C |
-30 to +50 °C |
Measuring range and accuracy correspond to probe |
0 to +50 °C |
Accuracy ±1 digit |
±0.5 °C |
±0.3 °C |
±(0.5 + 0.5 % of m.v.) °C |
±0.5 °C |
±0.5 °C |
|||
Resolution |
0.1 °C |
0.1 °C |
||||||
Humidity measurement |
||||||||
Measuring range |
– |
0 to 100 %RH |
Measuring range and accuracy correspond to probe |
0 to 100 %RH |
||||
Accuracy |
– |
±2 %RH |
±2 %RH at +25 °C and 20 to 80 %RH ±3 %RH at +25 °C and <20> >80 %RH ±1% RH hysteresis ±1% RH / year drift |
|||||
Resolution |
– |
0.1 %RH |
||||||
CO2 measurement |
||||||||
Measuring range |
– |
0 to 5 000 ppm |
||||||
Accuracy |
– |
±(50 ppm + 3 % of m.v.) at +25 °C Without external power supply: ±(100 ppm + 3 % of m.v.) at +25 °C |
||||||
Resolution |
– |
1 ppm |
||||||
Pressure measurement |
||||||||
Measuring range |
– |
600 to 1100 mbar |
||||||
Accuracy |
– |
±3 mbar at +22 °C |
||||||
Resolution |
– |
1 mbar |
||||||
Temperature/humidity measurement via external probes |
||||||||
External probe connection |
– |
External temperature probes |
External temperature probes |
– |
External humidity/ temperature probe |
– |
||
General technical data |
||||||||
Operating temperature |
-30 to +50 °C |
0 to +50 °C |
||||||
Storage temperature (without batteries) |
-40 to +70 °C |
-20 to +50 °C |
||||||
Protection class |
IP65 |
IP65 |
IP54 |
IP30 |
IP54 |
IP20 |
||
Measuring cycle |
Depends on the Cloud licence / Basic: 15 min to 24 h / Advanced: 1 min to 24 h testo 160 IAQ – Advanced in battery operation: 5 mins to 24 h |
|||||||
Communication cycle |
Depends on the Cloud licence / Basic: 15 min to 24 h / Advanced: 1 min to 24 h |
|||||||
Memory |
10000 readings/channel |
32,000 readings (sum of all channels) |
||||||
Standards / permits |
EN 12830 |
EN 12830 |
– |
|||||
Battery life |
12 months (typical value, depending on the wireless LAN infrastructure) at +25 °C, 15 mins measurement rate and standard communication rate at -30 °C, 15 mins measurement rate and standard communication rate with Energizer batteries 0515 0572 |
12 months |
||||||
Voltage supply |
4 x AA AlMn Mignon batteries; mains unit optional; for temperatures under -10 °C please use Energizer batteries 0515 0572 |
4 x AA alkali manganese batteries 1.5 V, alternatively mains unit via USB connection |
||||||
Dimensions |
95 x 75 x 30.5 mm |
115 x 82 x 31 mm |
95 x 75 x 30.5 mm |
117 x 82 x 32 mm |
||||
Weight (including batteries) |
240 g |
250 g |
240 g |
269 g |
||||
Door contact |
No |
Optional |
No |
|||||
Communication |
||||||||
WiFi (wireless LAN) |
Signal transmission: wireless; frequency band: 2.4 GHz; supported wireless LAN standards: IEEE 802.11 b/g/n and IEEE 802.1X Possible encryption methods: without encryption, WEP, WPA, WPA2, WPA2 Enterprise The data loggers communicate via the standard protocol MQTT and are SNTP time synchronization-capable. |
Accessories | Order no. |
Door contact for testo Saveris 2-T2 WiFi data logger | 0572 2152 |
Mains unit for testo Saveris 2 WiFi data loggers | 0572 2020 |
Battery for WiFi probes (4 x AA alkali manganese Mignon batteries)) | 0515 0414 |
Batteries for operating WiFi data loggers testo Saveris 2 below -10 °C (4 x Energizer L91 Photo-Lithium) | 0515 0572 |
Magnetic adapter for testo Saveris 2 wall bracket, for attachment to magnetic surfaces | 0554 2001 |
Wall bracket for testo 160 IAQ | 0554 2015 |
Deco-cover for testo 160 IAQ | 0554 2012 |
ISO temperature calibration certificate, temperature probe; calibration points -18 °C, 0 °C, +40 °C; per channel/instrument | 0520 0153 |
DAkkS temperature calibration certificate; temperature probe; calibration points -18 °C, 0 °C, +40 °C; per channel/instrument | 0520 0262 |
ISO humidity calibration certificate; calibration points 11.3 %RH and 75.3 %RH at +25 °C/ +77 °F; per channel/instrument | 0520 0076 |
DAkks humidity calibration certificate; humidity probe; calibration points 11.3 %RH and 75.3 %RH at +25°C; per channel/instrument | 0520 0246 |
ISO calibration certificate CO2, calibration points 0; 1000; 5000 ppm | 0520 0033 |