Talk with one of our sales

Monday to Friday 8AM - 5PM WIB

021- 72791867

Product

Home / Product

TESTO

Testo 160

Category: DATA LOGGER, TEMPERATURE

Model: Temperature, Temperature Data Logger

Avaibility: AVAILABLE

Call For Price

Monitoring system

testo 160 – Monitoring system for the monitoring of temperature, humidity,

light intensity, UV radiation and CO2 concentration.

The monitoring system testo 160 monitors ambient conditions in display cases, exhibition rooms and depots. The loggers transfer measurement values by wireless LAN to the online store (Testo Cloud). You can access all data   at any time with the testo Saveris 2 App or by PC/tablet/ smartphone and a normal browser. If limit values are exceeded, an alarm is immediately provided by SMS and/   or e-mail. For light intensity, an alarm can also be triggered  if the accumulated light quantity within a day, a week or a month exceeds a limit value.

Thanks to the optional, individually designable deco-cover, the loggers can be integrated inconspicuously in exhibitions and display cabinets. The temperature and humidity probe with wall bushing is ideal for monitoring small display cabinets in which a data logger cannot be placed.

The testo 160 thus enables you to check all relevant ambient conditions, in order to safeguard the value of the exhibits and fulfil the obligation of documentation.

 

The testo 160 Cloud

 

Basic

Advanced

Measuring cycle

15 min. to 24 h

1 min. to 24 h

Communication cycle

15 min. to 24 h

1 min. to 24 h

Data storage

Max. 3 months

Max. 2 years

Reports

Manual (.pdf/.csv)

Manual (.pdf/.csv) Automatic (.pdf/.csv)

Data analysis

For one measurement site each (external probes count as measurement sites)

For up to 10 measurement channels simultaneously

Number of users per account

1

10

Number of WiFi data loggers per account

Unlimited

Unlimited

 

Alarm options

 

Upper/lower alarm limits

  • Upper/lower alarm limits
    • Alarm delay
  • Time control of alarms

 

 

System notifications

 

  • Notification of low battery
    • Radio link interrupted
  • Power supply interrupted

 

  • Notification of low battery
    • Radio link interrupted
  • Power supply interrupted

E-mail alarm

Yes

Yes

SMS alarm

No

  • Including 25 SMS per logger and year
    • More SMS packages purchasable
 

12-month licence order no. 0526 0735

24-month licence order no. 0526 0732

36-month licence order no. 0526 0733

Technical data WiFi data loggers

 

WiFi

data logger testo 160 TH

WiFi

data logger testo 160 THE

WiFi

data logger testo 160 THL

WiFi

air quality logger testo 160 IAQ

WiFi

data logger testo 160 E

Temperature measurement

Measuring range

-10 to +50 °C

0 to +50 °C

see external probe

Accuracy

± 0.5 °C

Resolution

0.1 °C

Humidity measurement

Measuring range

0 to 100 %RH (non-condensing)

 

 

see external probe

 

Accuracy

±2 %RH at +25 °C and 20 to 80 %RH

±3 %RH at +25 °C and < 20> 80 %RH

±1 %RH hysteresis

±1 %RH / year drift

Resolution

0.1% RH

Lux measurement

Measuring range

 

 

 

see external probe

0 to 20,000 lux

 

 

 

see external probe

 

 

Accuracy

DIN 5032-7

Class C-compliant.

±3 lux or 3 % of m.v. (refers to reference DIN 5032-7 Class L)

Resolution

0.1 lux

UV measurement

Measuring range

 

 

 

see external probe

0 to 10,000 mW/m2

 

 

 

see external probe

 

Accuracy

±5 mW/m2 or

±5 % of m.v. (refers to external reference)

Resolution

0.1 mW/m2

CO2 measurement

Measuring range

 

0 to 5,000 ppm

 

 

 

Accuracy

±(50 ppm + 3 %  of m.v.) at +25 °C Without external

power supply: ±(100 ppm + 3 % of m.v.)

at +25 °C

Resolution

1 ppm

Pressure measurement

Measuring range

 

600 to 1100 mbar

 

Accuracy

±3 mbar

at  +22 °C

Resolution

1 mbar

WLAN

Standard

802.11 b/g/n

 

Security

WPA2 Enterprise: EAP-TLS, EAP-TTLS-TLS, EAP-TTLS-MSCHAPv2, EAP-TTLS-PSK, EAP-PEAP0-TLS, EAP-PEAP0-MSCHAPv2, EAP-PEAP0-PSK, EAP-PEAP1-TLS, EAP-PEAP1-MSCHAPv2, EAP-PEAP1-PSK, WPA

Personal, WPA2 (AES), WPA (TKIP), WEP

General

Operating temperature

-10 to +50 °C

0 to +50 °C

-10 to +50 °C

Storage temperature

-20 to +50 °C

0 to +50 °C

-20 to +50 °C

Protection class

IP20

Measuring cycle

Dependent on Cloud licence / Basic: 15 min to 24 h / Advanced: 1 min to 24 h testo 160 IAQ – Advanced in battery operation: 5 mins to 24 h

Communication cycle

Dependent on Cloud licence / Basic: 15 min to 24 h / Advanced: 1 min to 24 h

Memory

32,000 readings (sum of all channels)

Voltage supply (alternatively mains unit via USB connection)

 

4 x AAA alkaline manganese batteries 1.5 V

4 x AA alkaline man- ganese batteries

1.5 V

4 x AAA alkaline manganese batteries

1.5 V

Battery life

(depending on the measuring and communication cycle

vis-a-vis the Cloud)

 

1.5 years

 

1 year

 

1.5 years

Dimensions

76 x 64 x 22 mm

76 x 64 x 22 mm

92 x 64 x 22 mm

117 x 82 x 32 mm

76 x 64 x 22 mm

Weight

(including batteries)

94 g

94 g

113 g

269 g

96 g

 

         

 

Accessories

Order no.

   

Deco-cover for testo 160 TH / testo 160 THE / testo 160 E

0554 2006

Deco-cover for testo 160 THL

0554 2009

Deco-cover for testo 160 IAQ

0554 2012

Wall bracket for testo 160 TH / testo 160 THE / testo 160 E / testo 160 THL

0554 2013

Wall bracket for testo 160 IAQ

0554 2015

Extension cable for probes, length 0.6 m (included with every probe)

0554 2004

Extension cable for probes, length 2.5 m

0554 2005

Display cabinet bushing for temperature and humidity probes (included with every probe)

0554 2016

Alkaline manganese microcell AAA batteries up to -10 °C, order 4 off

0515 0009

Alkaline manganese mignoncell AA batteries up to -10 °C, order 4 off

0515 0414

External USB power supply

0572 2020

ISO calibration certificate temperature -8 °C, 0 °C, +40 °C (for testo 160 TH / testo 160 THE / testo 160 E / testo 160 THL)

0520 0171

ISO calibration certificate temperature +15 °C, +25 °C, +35 °C (für testo 160 IAQ)

0520 0172

ISO calibration certificate humidity at +25 °C, humidity points 11.3 %RH and 75.3 %RH

0520 0076

ISO calibration certificate light intensity, calibration points 0; 500; 1000; 2000; 4000 Lux

0520 0010

ISO calibration certificate CO2, calibration points 0; 1000; 5000 ppm

0520 0033

 

OTHER PRODUCT

We Offer Many Product For All Your Needs

Need help buying the right product?

We’ll assist you for further information, application assistance, or to schedule a technical meeting.