BEAMEX

Beamex FB -- Portable Dry-Block Calibrators

Category: TEMPERATURE, CALIBRATORS

Model: Dry Block Calibrator , Dry Block Calibrators

Avaibility: AVAILABLE

Call For Price

Lightweight, portable

The Beamex FB field temperature block is ideal for industrial field use. It only weighs about 8 kg, and it is small enough to carry around.

Speed

The Beamex FB is extremely quick to reach various temperatures, i.e. it cools down to –25 °C in 15 minutes and heats up to +660 °C in 15 minutes. This saves time and increases productivity.

Accuracy and performance

The Beamex FB is an easily portable unit that also provides excellent calibration accuracy. The display accuracy is up to ±0.2 °C and its control technology provides great stability up to ±0.01 °C. The dual zone controlled block provides excellent axial uniformity up to ±0.04 °C and radial uniformity up to ±0.01 °C.

Smart reference sensors

The Beamex FB has an internal reference thermometer (in R models), which enables connections to the Beamex smart reference sensors. These sensors have a memory that contains all of the sensor correction data. This enables the use of the reference sensor as a real plug-and-play.

Accredited calibration

Each Beamex FB field temperature block is delivered with an accredited calibration certificate.

Usability

The large LCD display, function keys and multilingual, menu-based user interface makes the Beamex FB easy to use. A graphic and audible stability indicator lets you know when a block is stable. The HOT warning light indicates when the block is hot (over +50 °C). It blinks as long as the block is too hot to touch, even when the unit is switched off or when the mains cable is disconnected.

Part of the ICS integrated calibration solution

The communication port enables communication with selected Beamex MC calibrators for automation calibration and documentation, making the Beamex FB products part of the Beamex ICS integrated calibration solution. Combined with the Beamex MC6 calibrator, loop calibrations are possible with conventional, HART and Fieldbus temperature transmitters with sensors.

 

FB150

FB350

FB660

Temperature range at 23 °C

–25 °C to 150 °C (–13 °F to 302 °F)

33 °C to 350 °C (91 °F to 662 °F)

50 °C to 660 °C (122 °F to 1220 °F)

Display accuracy

+/-0.2 °C Full range

+/-0.2 °C Full range

+/-0.35 °C at 50 °C +/-0.35 °C at 420 °C +/-0.5 °C at 660 °C

Stability

+/-0.01 °C Full range

+/-0.02 °C at 33 °C +/-0.02 °C at 200 °C +/-0.03 °C at 350 °C

+/-0.03 °C at 50 °C +/-0.05 °C at 420 °C +/-0.05 °C at 660 °C

Axial uniformity at 40 mm (1.6 in)

+/-0.05 °C Full range

+/-0.04 °C at 33 °C +/-0.1 °C at 200 °C +/-0.2 °C at 350 °C

+/-0.05 °C at 50 °C +/-0.35 °C at 420 °C +/-0.5 °C at 660 °C

Radial uniformity

+/-0.01 °C Full range

+/-0.01 °C at 33 °C +/-0.015 °C at 200 °C +/-0.02 °C at 350 °C

+/-0.02 °C at 50 °C +/-0.05 °C at 420 °C +/-0.10 °C at 660 °C

Loading effect
(with a 6.35 mm reference probe and three 6.35 mm probes)

+/- 0.006 °C Full range

+/- 0.015 °C Full range

+/- 0.015 °C at 50 °C +/- 0.025 °C at 420 °C +/- 0.035 °C at 660 °C

Hysteresis

+/-0.025 °C

+/-0.03 °C

+/-0.01 °C

Immersion depth

150 mm (5.9 in)

Insert OD dimensions

30 mm (1.18 in)

25.3 mm (0.996 in)

24.4 mm (0.96 in)

Heating time

16 min: 23 °C to 140 °C 23 min: 23 °C to 150 °C 25 min: –25 °C to 150 °C

5 min: 33 °C to 350 °C

15 min: 50 °C to 660 °C

Cooling time

15 min: 23 °C to –25 °C 25 min: 150 °C to –25 °C

32 min: 350 °C to 33 °C 14 min: 350 °C to 100 °C

35 min: 660 °C to 50 °C 25 min: 660 °C to 100 °C

Resolution

0.01 °C / °F

Display

LCD, °C or °F user-selectable

Size (H x W x D)

290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)

Weight

8.16 kg (18 lb)

7.3 kg (16 lb)

7.7 kg (17 lb)

Power requirements

230 V (+/-10%) 50/60 Hz, 575 W
100 V to 115 V (+/-10%) 50/60 Hz, 635 W

230 V (+/-10%), 50/60 Hz, 1800 W 100 V to 115 V (+/-10%), 50/60 Hz, 1400 W

230 V (+/-10%), 50/60 Hz, 1800 W 100 V to 115 V (+/-10%), 50/60 Hz, 1400 W

Computer interface

RS-232

RS-232

RS-232

Calibration

Accredited calibration certificate provided

Environmental operating conditions

0 °C to 50 °C, 0% to 90% RH (non-condensing)

Specifications valid in environmental conditions

13 °C...33 °C