Beamex MB -- Portable Dry-Block calibrators
Category: TEMPERATURE, CALIBRATORS
Model: Dry Block Calibrator , Dry Block Calibrators
Avaibility: AVAILABLE
Call For Price
Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
Beamex MB -- Portable Dry-Block calibrators
Category: TEMPERATURE, CALIBRATORS
Model: Dry Block Calibrator , Dry Block Calibrators
Avaibility: AVAILABLE
Call For Price
High accuracy and stability
With a conventional dry block, you typically needed an external reference sensor if you wanted better accuracy. The Beamex MB has accurate internal temperature measurement and display accuracy up to ±0.1°C, so you can get high accuracy even without an external reference sensor. With the unique temperature control techniques the Beamex® MB has excellent stability up to ±0.005°C. This kind of stability has usually been found only in baths, not in dry blocks.
Built-in high-accuracy reference input
In order to receive the best accuracy from the MB, there is a possibility to connect an external reference sensor into the reference sensor connection (R model). This eliminates the need for a separate reference thermometer. The reference sensor measurement is accurate up to ±0.006 °C. ITS-90 or CVD coefficients can be used to compensate any sensor errors.
Axial uniformity
With the unique dual zone control and extended well depth, the Beamex MB has an excellent axial uniformity up to ±0.02 °C.
Radial uniformity
Radial uniformity is the temperature difference between the holes in the insert. It is naturally crucial that the reference sensor and the sensor being tested are at the same temperature. The Beamex MB offers radial uniformity up to ±0.01 °C.
Immersion depth
The Beamex MB series provides immersion depth up to 203 mm (160 mm in MB140), which, together with the control techniques, provides more stable calibration. Moreover, a deeper immersion depth reduces the stem conduction error (heat loss into the atmosphere), especially in higher temperatures.
Loading
With the extended well depth and the dual zone temperature control feature, the Beamex MB can correct the effect of loading and provides loading specifications up to ±0.005 °C.
MB140 |
MB155 |
MB425 |
MB700 |
|
Temperature range at 23 °C |
–45 °C to 140 °C (–49 °F to 284 °F) |
–30 °C to 155 °C (–22 °F to 311 °F) |
35 °C to 425 °C (95 °F to 797 °F) |
50 °C to 700 °C3) (122 °F to 1292 °F) |
Display accuracy |
±0.1 °C Full range |
±0.1 °C Full range |
±0.1 °C to 100 °C ±0.15 °C to 225 °C ±0.2 °C to 425 °C |
±0.2 °C to 425 °C ±0.25 °C to 660 °C |
Stability |
±0.005 °C Full range |
±0.005 °C Full range |
±0.005 °C to 100 °C ±0.008 °C to 225 °C ±0.01 °C to 425 °C |
±0.005 °C to 100 °C ±0.01 °C to 425 °C ±0.03 °C to 700 °C |
Axial uniformity 40 mm (1.6 in) |
±0.08 °C to –35 °C ±0.04 °C to 0 °C ±0.02 °C to 50 °C ±0.07 °C to 140 °C |
±0.025 °C to 0 °C ±0.02 °C to 50 °C ±0.05 °C to 155 °C |
±0.05 °C to 100 °C ±0.09 °C to 225 °C ±0.17 °C to 425 °C |
±0.09 °C to 100 °C ±0.22 °C to 425 °C ±0.35 °C to 700 °C |
Radial uniformity |
±0.01 °C Full range |
±0.01 °C Full range |
±0.01 °C to 100 °C ±0.02 °C to 225 °C ±0.025 °C to 425 °C |
±0.01 °C to 100 °C ±0.025 °C to 425 °C ±0.04 °C to 700 °C |
Loading effect (with a 6.35 mm reference probe and three 6.35 mm probes) |
±0.02 °C to –35 °C ±0.005 °C to 100 °C ±0.01 °C to 140 °C |
±0.005 °C to 0 °C ±0.005 °C to 100 °C ±0.01 °C to 155 °C |
±0.01 °C Full range |
±0.02 °C to 425 °C ±0.04 °C to 700 °C |
Hysteresis |
±0.025 °C |
±0.025 °C |
±0.04 °C |
±0.07 °C |
Immersion depth |
160 mm (6.3 in) |
203 mm (8 in) |
203 mm (8 in) |
203 mm (8 in) |
Resolution |
0.001 °C / °F |
|
|
|
Display |
LCD, °C or °F, user-selectable |
|
|
|
Key pad Ten key with decimal and +/– button. Function keys, menu key, and °C / °F key. |
||||
Insert OD dimensions |
30.0 mm (1.18 in) |
30.0 mm (1.18 in) |
30.0 mm (1.18 in) |
29.2 mm (1.15 in) |
Cooling time |
44 min: 23 °C to –45 °C 19 min: 23 °C to –30 °C 19 min: 140 °C to 23 °C |
30 min: 23 °C to –30 °C 25 min: 155 °C to 23 °C |
220 min: 425 °C to 35 °C 100 min: 425 °C to 100 °C |
235 min: 700 °C to 50 °C 153 min: 700 °C to 100 °C |
Heating time |
32 min: 23 °C to 140 °C 45 min: –45 °C to 140 °C |
44 min: 23 °C to 155 °C 56 min: –30 °C to 155 °C |
27 min: 35 °C to 425 °C |
46 min: 50 °C to 700 °C |
Size (H x W x D) |
366 x 203 x 323 mm (14.4 x 8 x 12.7 in) |
|
|
|
Weight |
14.2 kg (31.5 lb) |
14.6 kg (32 lb) |
12.2 kg (27 lb) |
14.2 kg (31.5 lb) |
Power requirements |
230 VAC (±10%), 550 W 115 VAC (±10%), 550 W |
230 VAC (±10%), 550 W 115 VAC (±10%), 550 W |
230 VAC (±10%), 1025 W 115 VAC (±10%), 1025 W |
230 VAC (±10%), 1025 W 115 VAC (±10%), 1025 W |
Computer interface |
RS-232 |
|
|
|
Calibration Accredited calibration certificate provided |
||||
Environmental operating 5 °C to 40 °C, 0% to 80% RH (non-condensing) conditions |
||||
Specifications valid in environmental conditions |
18 °C…28 °C |
|
|
|
OTHER PRODUCT