With its simple software, long battery life and large memory, the GP1 is very easy to set up and maintain. It is well suited to outdoor logging applications ? the waterproof (IP67) case does not require a secondary enclosure. The GP1's small size allows it to be installed in confined spaces, or to be hidden away if security is an issue. The internal alkaline battery lasts for >1 year when taking hourly readings from 2 moisture sensors, 2 temperature sensors and a rain gauge.
- High accuracy 7 channel data logger
- 600,000 readings
- Compatible with DeltaLINK?Cloud online data viewing and sharing platform
- Smart relay control
Applications
? Monitoring soil moisture
? Environmental data logging
? Climate research and meteorolog
|
Specifications |
Typical at +20 °C |
Max ?20 °C to +60 °C |
Notes |
|
|
Differential Voltage Channels |
Voltage accuracy |
± (0.3 mV + 0.01% reading) |
± (1.6 mV + 0.05% reading) |
over full ?0.2 V to +2.7 V voltage range |
|
Soil moisture accuracy |
±0.06%? (±0.0006 m3.m?3) |
±0.3%vol? (±0.003 m3.m?3) |
with ML3 or SM150T (0 to 60% ?) [1] |
|
|
Resolution / input noise |
±0.1 mV |
effective resolution of readings (typical) |
||
|
Input voltage range |
?0.2 V to +2.7 V |
|
||
|
Input voltage limits |
?2.8 V to +3.6 V [2] |
each input signal relative to logger GND |
||
|
Temperature Channels |
Temperature accuracy |
±0.07 °C |
±0.1 °C (with GP1 below 30 °C if reading ?20 to 0 °C) |
using 10K thermistor ?20 to +60 °C [1] |
|
Resistance accuracy |
±0.2% of 2 to 100K? reading (±0.3% max) |
±0.5% of 2 to 20K? reading |
|
|
|
Event Counters |
Event counter (Event6) |
< 50> |
logic low input < 1> 1.9 V, maximum ±14 V on Event inputs 5 and 6 |
|
|
High speed counter (Event5) |
< 33k> |
|||
|
Power |
Internal battery life |
1 year typical (alkaline) |
9V PP3 battery [3] |
|
|
External power |
11 to 24 V DC |
power via external M8 connector |
||
|
Switched sensor power |
up to 120 mA: > 10 V if external power > 13.8 V |
switched battery or external power |
||
|
+5 V reference |
5 V ±0.6% |
5 V ±1.6% |
switched voltage reference, up to 50 mA |
|
|
Relay Channel |
1 x relay |
SPST, < 30> |
separate ON / OFF conditions adjustable ON duty cycle |
|
|
Data Recording |
Logging frequency |
1 second to 24 hour |
user configurable logging frequency |
|
|
Sensor warm?up |
multiples of 1 second |
user configurable sensor warm?up times |
||
|
Internal flash memory |
> 600k readings, typical |
non?volatile flash memory |
||
|
Communications |
RS232 (115 kbaud) / USB [4] |
comms via external M8 connector or USB [4] |
||
|
Physical |
Environmental |
water resistant to IP67 |
4 cable glands, connector & case |
|
|
Size and weight |
140 x 105 x 45 mm, 280 g |
including battery |
||
|
Temperature |
?20 to +60 °C |
|
||