Fluke 726 PRECISION MULTIFUNCTION CALIBRATORS
Category: CALIBRATORS
Model: Multifunction Calibrators
Avaibility: AVAILABLE
Call For Price
-10.00 mV to +75.00 mV
726 Specifications
| Function Measure or Source | Range or Type | Resolution | Accuracy | Notes |
| Voltage | 0 to 100 mV 0 to 20 V (source) 0 to 30 V (measure) | 0.01 mV 0.01 V 0.01 V | 0.01 % Rdg + 2 LSD | Max load, 1 mA |
| mA | 0 to 24 | 0.001 mA | 0.01 % Rdg + 2 LSD | Max load, 1000 ? |
| mV (TC terminals) | -10.00 mV to +75.00 mV | .01 mV | 0.01 % of range + 1 LSD | |
| Resistance | 5 ? to 4000 ? | ? 0.01 ? to 0.1 ? | 0.015 % | |
| Frequency | 2.0 to 1,000 CPM 1 to 1100 Hz 1.0 to 10.0 kHz 10.0 to 15.0 kHz | 0.1 CPM 1 Hz 0.1 kHz 0.1 kHz | ± 0.05 % ± 0.05 % ± 0.25 % ± 0.5 % | For frequency source; 1 V to 20 V p-p squarewave, -0.1 V offset |
| Loop Supply | 24 V dc | N/A | 10 % | |
| Thermocouples | J, K, T, E, L, N, U, C, BP, XK | 0.1 °C, 0.1 °F | to 0.2 °C | |
| Thermocouples | B, R, S | 1 °C, 1 °F | to 1.2 °C | |
| RTDs | Cu 10 Ni 120 (672) Pt 100, 200, 500, 1000 (385) Pt 100 (3916), Pt 100 (3926) Cu 10 | 0.1 °C, 0.1 °F 0.01 °C, 0.01 °F | to 1.8 °C to 0.15 °C |
| Simultaneous Function Capability | Channel A | Channel B |
| 24.000 mA DC | M | M or S |
| 24.000 mA DC with 24 V loop supply | M | |
| 100.00 mV DC | M or S | |
| 30.000 V DC measure | M | |
| 20.000 V DC measure, 20.000 VDC Source | M or S | |
| 5 to 4000 Ohms | M or S | |
| Thermocouple J, K, T, E, R, S, B, M, L, U, N, C, BP, XK | M or S | |
| RTD Ni120; Pt100 (392); Pt100 (JIS); Pt100, 200, 500, 1000 (385), Cu 10 | M or S | |
| Pressure (requires Fluke 700PXX Modules) | M | M used as S |
| Frequency; Squarewave, 1 CPM to 10 kHz; fixed amplitude 5 V p-p | M or S |
M = Measure, S = Source/Simulate