Fluke 1524 HANDHELD THERMOMETER READOUT
Category: TEMPERATURE
Model: Handheld Temperature Calibrator and Indicator
Avaibility: AVAILABLE
Call For Price
Talk with one of our sales
Monday to Friday 8AM - 5PM WIB
021- 72791867
Fluke 1524 HANDHELD THERMOMETER READOUT
Category: TEMPERATURE
Model: Handheld Temperature Calibrator and Indicator
Avaibility: AVAILABLE
Call For Price
PRTs and thermistors: 0.001°
Specifications
| Input channels | 2 |
| Resolution | PRTs and thermistors: 0.001° thermocouples: 0.01° |
| Logging | 25 readings with statistics15,000 time and date stamped |
| Sample interval (normal) | 1 second (simultaneous measurement) |
| Typical sample interval (fast mode)* | 0.3 seconds |
| Sensor types | PRTs, RTDs, thermistors, and thermocouples |
| Statistics | Maximum, minimum, average, standard deviation |
| Trending | Scale: ± 10 °C (18 °F), ± 1 °C (1.8 °F), ± 0.1°C (0.18 °F), ± 0.01°C (0.018 °F), 10 minutes of real-time data |
| Power requirements | 3 AA alkaline batteries, 12 V dc universal power supply |
| Size (HxWxD) | 96 mm x 200 mm x 47 mm (3.75 in x 7.9 in x 1.86 in) |
| Weight | 0.65 kg (1.4 lb) |
| Computer interface | RS-232, 9940 I/O ToolKit software included |
| Safety | EN61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 |
| Environmental conditions for best accuracy: 13 °C to 33 °C (55.4 °F to 91.4 °F) | |
| Millivolt range and accuracy | -10 mV to 75 mV, ± (0.005 % + 5 µV) |
| Internal reference junction compensation | ± 0.2 °C (± 0.36 °F) |
| Resistance range and accuracy | 0 ? to 400 ? ± (0.004 % + 0.002 ?) 200 ? to 500 k? ± (0.01 % + 0.5 ?) 50k? to 500 k? ± (0.03 %) |
| Temperature coefficient, voltage: -10 °C to 13 °C, +33 °C to 60 °C (14 °F to 55.4 °F, 91.4 °F to 140 °F) | ± (0.001 %/°C + 1 mV/°C) |
| Temperature coefficient, resistance: -10 °C to 13 °C, +33 °C to 60 °C (14 °F to 55.4 °F, 91.4 °F to 140 °F) | 0.0008 %/°C + 0.0004 ? (0 ? to 400 ?) 0.002 %/°C + 0.1 ? (0 ? to 50 k?) 0.06 %/°C + 0.1 ? (50 k? to 500 k?) |
| Excitation current, resistance | 1 mA (0 ? to 400 ?) 10 µA (0 ? to 50 k?) 2 µA (50 k? to 500 k?) |
*See technical manual for sample interval details by sensor type and number of inputs.