Fluke 9142, 9143, & 9144 Field Metrology Wells Specifications

Fluke 9142, 9143, & 9144 Field Metrology Wells Specifications

Base Unit Specifications

Parameter914291439144
Temperature Range at 23 °C-25 °C to 150 °C (-13 °F to 302 °F)33 °C to 350 °C (91 °F to 662 °F)50 °C to 660 °C (122 °F to 1220 °F)
Display Accuracy± 0.2 °C Full Range± 0.2 °C Full Range± 0.35 °C at 50 °C ± 0.35 °C at 420 °C ± 0.5 °C at 660 °C
Stability± 0.01 °C Full Range± 0.02 °C at 33 °C ± 0.02 °C at 200 °C ± 0.03 °C at 350 °C± 0.03 °C at 50 °C ± 0.04 °C at 420 °C ± 0.05 °C at 660 °C
Axial Uniformity at 40 mm (1.6 in)± 0.05 °C Full Range± 0.05 °C at 33 °C ± 0.1 °C at 200 °C ± 0.2 °C at 350 °C± 0.05 °C at 50 °C ± 0.2 °C at 420 °C ± 0.3 °C at 660 °C
Radial Uniformity± 0.01 °C Full Range± 0.01 °C at 33 °C ± 0.015 °C at 200 °C ± 0.02 °C at 350 °C± 0.02 °C at 50 °C ± 0.05 °C at 420 °C ± 0.14 °C at 660 °C
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes)± 0.006 °C Full Range± 0.015 °C Full Range± 0.015 °C at 50 °C ± 0.025 °C at 420 °C ± 0.035 °C at 660 °C
Hysteresis0.0250.030.1
Operating Conditions0 °C to 50 °C, 0 % to 90 % RH (non-condensing)0 °C to 50 °C, 0 % to 90 % RH (non-condensing)0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
Environmental Conditions (for all specifications except temperature range)13 °C to 33 °C13 °C to 33 °C13 °C to 33 °C
Immersion (Well) Depth150 mm (5.9 in)150 mm (5.9 in)150 mm (5.9 in)
Insert OD30 mm (1.18 in)25.3 mm (1.00 in)24.4 mm (0.96 in)
Heating Time16 min: 23 °C to 140 °C 23 min: 23 °C to 150 °C 25 min: -25 °C to 150 °C5 min: 33 °C to 350 °C15 min: 23 °C to 660 °C
Cooling Time15 min: 23 °C to -25 °C 25 min: 150 °C to -23 °C32 min: 350 °C to 33 °C 14 min: 350 °C to 100 °C35 min: 660 °C to 50 °C 25 min: 660 °C to 100 °C
Resolution0.01 °0.01 °0.01 °
DisplayLCD, °C or °F user-selectableLCD, °C or °F user-selectableLCD, °C or °F user-selectable
Size (H x W x D)290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
Weight8.16 kg (18 lb)7.3 kg (16 lb)7.7 kg (17 lb)
Power Requirements100 V to 115 V (± 10 %) 50/60 Hz, 632 W 230 V (± 10 %) 50/60 Hz, 575 W100 V to 115 V (± 10 %), 50/60 Hz, 1380 W 230 V (± 10 %), 50/60 Hz, 1380 W100 V to 115 V (± 10 %), 50/60 Hz, 1380 W 230 V (± 10 %), 50/60 Hz, 1380 W
Computer InterfaceRS-232 and 9930 Interface-it control software includedRS-232 and 9930 Interface-it control software includedRS-232 and 9930 Interface-it control software included

-P Specifications (Process Option)

Built-in Reference Thermometer Readout Accuracy (4-Wire Reference Probe)¹± 0.010 °C at -25 °C ± 0.015 °C at 0 °C ± 0.020 °C at 50 °C ± 0.025 °C at 150 °C ± 0.030 °C at 200 °C ± 0.040 °C at 350 °C ± 0.050 °C at 420 °C ± 0.070 °C at 660 °C
Reference Resistance Range0 ohms to 400 ohms
Reference Resistance Accuracy²0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±60 ppm of reading
Reference CharacterizationsITS-90, CVD, IEC-751, Resistance
Reference Measurement Capability4-wire
Reference Probe Connection6-pin Din with Infocon Technology
Built-in RTD Thermometer Readout AccuracyNI-120: ± 0.1 °C at 0 °C PT-100 (385): ± 0.02 °C at 0 °C PT-100 (3926): ± 0.02 °C at 0 °C PT-100 (JIS): ± 0.02 °C at 0 °C
RTD Resistance Range0 ohms to 400 ohms
RTD Resistance Accuracy²0 ohms to 25 ohms: ±0.002 ohms 25 ohms to 400 ohms: ±80 ppm of reading
RTD CharacterizationsPT-100 (385), (JIS), (3926), NI-120, Resistance
RTD Measurement Capability4-wire RTD (2-, 3-wire RTD w/ jumpers only)
RTD Connection4 terminal input
Built-in TC Thermometer Readout AccuracyType J: ± 0.7 °C at 660 °C Type K: ± 0.8 °C at 660 °C Type T: ± 0.8 °C at 400 °C Type E: ± 0.7 °C at 660 °C Type R: ± 1.4 °C at 660 °C Type S: ± 1.5 °C at 660 °C Type M: ± 1.4 °C at 660 °C Type L: ± 0.7 °C at 660 °C Type U: ± 0.7 °C at 660 °C Type N: ± 0.9 °C at 660 °C Type C: ± 1.1 °C at 660 °C
TC Millivolt Range-10 mV to 75 mV
Voltage Accuracy-10 mV to 50 mV: ± 0.01 mV 50 mV to 75 mV: ± 250 ppm of reading
Internal Cold Junction Compensation Accuracy± 0.5 °C (ambient of 13 °C to 33 °C)
TC ConnectionSmall connectors
Built-in mA Readout Accuracy0.02 % of reading + 2 µV
mA RangeCal 4–22 mA, Spec 4–24 mA
mA Connection2 terminal input
Loop Power Function24–28 V dc loop power
Built-in Electronics Temperature Coefficient (0 °C to 13 °C, 33 °C to 50 °C)± 0.005 % of range per °C

¹ The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors.

² Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.